The goal of this blog is to become a dynamical way of providing fast and updated useful information, to the user of the instrument OSIRIS at GTC, complementing the information provided by GRANTECAN S.A.
It contains information about the manuals, instrument calibration, pipelines for download, tutorials, contributions, announcements, etc. Also, some historical information about the integration, and commissioning of the different observing modes is provided as well.
This Blog is maintained by the OSIRIS Scientific Team.

Saturday, July 29, 2006

Pruebas en frío - CCDs (I)







Operaciones de enfriado del criostato de OSIRIS con CCDs.
En las fotografías, consolas de control de temperatura y vacío del dispositivo, y primeras imágenes de un CCD-grado5 (parcialmente operativo, con columnas de píxeles defectuosas) a -90ºC.

Thursday, July 20, 2006

Ensamblando el Criostato (2) - CCDs








Se están llevando a cabo las pruebas criogénicas con CCDs.
En las fotografías, los sucesivos pasos del montaje del mosaico formado por dos CCDs tipo E2V44-82. Los test preliminares se realizan con un detector grado 6 (dummy) y uno de grado 5 (también llamado de ingeniería, parcialmente operativo).
La temperatura de trabajo es de 183.2 K (= -90ºC), y el bucle de control está conducido por un equipo Lakeshore 331, junto con sensores de temperatura (diodos de Silicio) y resistencias calentadoras. La potencia necesaria para que el servo PID estabilice la temperatura de trabajo es de aproximadamente 2 Wats.